Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2300150241127205412

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Fastcal stops and then the software shuts down See Long Description for details Fastcal stops and then the software shuts down with Error: 230015024 A corrupt file has been found in the cal database Error: 230015013 Failure opening a file for reading in the Cal DatabaseError: 230015035 Cal_db failure, reason: Address already in use gesyslog shows the following: Tue Sep 20 08:36:51 2005 Error: 230015024 Host: ls16 Process: scanmgr File: cal_db_file_io.c Line: 1983 Function: CAL Processing : SBC Processing A corrupt file has been found in the cal database . The corrupt file has been removed from the cal database and saved for later analysis. Corrupt File: /usr/g/cal_db/kv120/large_sfov/16x0625/large_spot/dasgain5/cal.1.acal Saved File: /usr/g/cal_db/recovery/kv120.large_sfov.16x0625.large_spot.das.cal.1.acal.1127205412 Tue Sep 20 08:36:51 2005 Error: 230015013 Host: ls16 Process: scanmgr File: cal_db_file_io.c Line: 226 Function: CAL Processing : SBC Processing Failure opening a file for reading in the Cal Database. File: /usr/g/cal_db/kv120/large_sfov/16x0625/large_spot/dasgain5/cal.1.sin Tue Sep 20 08:36:51 2005 Error: 230015035 Host: ls16 Process: scanmgr File: cal_db_file_io.c Line: 231 Cal_db failure, error number = 98, reason: Address already in use

Error message

Error: 230015024 A corrupt file has been found in the cal database See LD for complete error Tue Sep 20 08:36:51 2005 Error: 230015024 Host: ls16 Process: scanmgr File: cal_db_file_io.c Line: 1983 Function: CAL Processing : SBC Processing A corrupt file has been found in the cal database. The corrupt file has been removed from the cal database and saved for later analysis. Corrupt File: /usr/g/cal_db/kv120/large_sfov/16x0625/large_spot/dasgain5/cal.1.acal Saved File: /usr/g/cal_db/recovery/kv120.large_sfov.16x0625.large_spot.das.cal.1.acal.1127205412

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues