GE Healthcare · CTLightSpeed2601120225021926011400410132
LightSpeed 5.X (80/100kW)
Symptom
START button at the SCIM does not flash after confirming the scan
Error message
Error: 260112022 Communication between the TGP and Console Pushbuttons (SCIM) are down See long description for full error Wed Sep 10 00:06:08 2008 Error: 260112022 Host: tgp Process: tgp_subsys File: uif_table_tilt_mgr_gt.c Line: 5089 Wed Sep 10 00:06:08 2008 Host : Table Gantry Processor Ermes # : 260112022 Exception Class : Secondary Severity : Pri/Soft File : uif_table_tilt_mgr_gt.c Line# : 5089 Function : No System Function Reported Scan Type : Axial Scan: 50219/1/1 Communication between the TGP and Console Pushbuttons (SCIM) are down Error: 260114004 DUART Errors with the Console Pushbuttons. See Long Description for complete error. Wed Jan 24 10:39:46 2007 Error: 260114004 Host: tgp Process: tgp_subsys File: cpb_duart_isr.c Line: 62 Wed Jan 24 10:39:47 2007 Host : Table Gantry Processor Ermes # : 260114004 Exception Class : Secondary Severity : Pri/Soft File : cpb_duart_isr.c Line# : 62 Function : No System Function Reported Scan Type : None/Unknown Scan: 10132/2/0 DUART Errors with the Console Pushbuttons. HW Error: Possible hardware problem.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2404320 CONSOLE INTERCOM PRESCRIBED-TILT HUB See Long description for details Part #: 2404320 Part Type: GP CONSUMABLES Part Weight: 21.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: CONSOLE INTERCOM PRESCRIBED-TILT HUB FRU: Yes Status: Orderable Comments: ** FE ASSIST ** CT/Pet Discovery systems must use 2404320. Recommended for CT Systems to order 2404320-2 , to avoid gantry lights being displayed when console is shut down. FMI 25407 or 22918 must be loaded prior to using the 2404320-2. (212001261 10/21/08) Added per VCT RSPL, ms13452 041209 __Set as Critical Part for Product CT_VCT, ms04278, 05Oct14
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.