GE Healthcare · CTLightSpeed26000738650056260007422
LightSpeed 5.X (60kW)
Symptom
Scanner Hardware Stopped Scan with DAS parity errors
Error message
Error: 260007386 DAS Has Detected An Error. Data Parity Error detected in the xx Data Stream(s) on Converter Board xx or the DCB Board. See LD for sample errors Wed Mar 23 15:46:11 2005 Error: 260007386 Host: DAS Process: DCB Firmware File: parity.c Line: 219 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 214/5/1 Type: Helical Exception Level: Sec/Soft Time: 15:46:11:545 Log Series: 660 SCAN ABORT -- DAS Has Detected An Error. Data Parity Error detected in the 1B Data Stream (s) on Converter Board 33 or on Converter Board 41 . NOTE: Before troubleshooting, go through gesys.log (viewlog) and search for all “Data Parity Error” reports in the log. There are 2 items you need to note from the error: 1- Data Parity Error detected in the xx Data Stream and 2- On Converter Board xx or on Converter Board xx Tue Jan 10 09:37:14 2006 Error: 260007386 Host: orp Process: orp_subsys File: parity_16.c Line: 255 Exception Class : Abort Severity : Secondary Function : No System Function Reported Scan Type : Static Scan: 50056/1/1 SCAN ABORT -- DAS Has Detected An Error. Data Parity Error detected in the 3A Data Stream (s) on Converter Board 55 or on Converter Board 71 . Note: Error 260007
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2338891 RIBBON CABLE - MDAS SIGNAL (H16) 2338891 CABLE - DAS SIGNAL (H16) 2338891 RIBBON CABLE - MDAS SIGNAL (H16) Part #: 2338891 Part Type: GP CONSUMABLES Part Weight: 0.64 LBR Repairable: No Returnable: No Hazardous Material: N Description: CABLE - DAS SIGNAL (H16) FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required