Symptom

Unable to scan with a message on screen that says: Collimator/DAS reset failed.

Error message

Error: 260004025 DAS / Collimator firmware object: Lost communications with node applications. See Long Description for full error Thu Jan 13 06:58:23 2005 Error: 260004025 Host: OBC Process: Collimator File: col_cf_in_mail.c Line: 442 Function: No System Function Reported Scan: 50229/1/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 06:58:23:235 Log Series: 47 Collimator firmware object: Lost communications with node applications. Thu Dec 31 11:45:03 2009 Error:260004025 Host: OBC Process: DAS File: das cf in mail.c Line: 430 Function: No System Function Reported Scan: 19127/2/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 11:45:03:305 Log Series: 451 DAS firmware object: Lost communications with node applications.

Do Not Use - use one of the other Error: 260001008 The OBC failed to connect with the ? (DCB and/or the CCB) sub-system in the allotted time Open Long Description to see error example Mon Mar 28 03:39:48 2005 Error: 260001008 Host: OBC Process: HOP File: hop_node_init.c Line: 120 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 03:39:46:170 Log Series: 31 The OBC failed to connect with the DCB

Applies to

Models
LightSpeed
Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2159578-3 collimator control board used on 1X/2X/3X/4X/5X See Long description for details Part #: 2159578-3 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: COLLIMATOR CONTROL BOARD FRU: Yes Status: Orderable Comments: Not Available
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.