GE Healthcare · CTLightSpeedVCT29001000529000901229000900931809290010002
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Image recon is failing intermittently - document in RFS - critical issue See long description for details. Image recon is failing intermittently.
Error message
Error: 290010005 Socket error: socket is terminated because IG dies (error text varies) See Long Description for error examples. Tue Mar 16 10:40:09 2004 290010005 darc data_restore request_object.cxx 503 Socket error: socket is terminated because IG dies Fri Nov 16 19:22:57 2007 Error: 290010005 Host: darc Process: data_restore File: request_object.cxx Line: 556 Socket error: socket event from IG Error: 290009012 Socket event on 1, fd: 46. See long description for sample error message. Thu Feb 21 07:57:21 2008 Error: 290009012 Host: darc Process: recon_control File: IGControl.cpp Line: 1069 Socket event on 1, fd: 46. Do not Use - use one of the other Error: 290009009 Image failed due to many reasons See Long Description for complete error Tue Apr 19 12:44:49 2005 Error: 290009009 Host: darc Process: recon_control File: ImageSetRequest.cxx Line: 3968 Function: Image Generation : SBC Processing Job Type: RECON_JOB_PROS_AXIAL_RECON Scan: 4354/2/3 Image: 4354/2/6 Image failed due to: IG error: PROCESSING_ERROR. Number of retry so far: 1. Tue Nov 7 11:38:44 2006 Error: 290009009 Host: darc Process: recon control File: ImageSetRequest.cxx Line: 5348 Function: Image Generation : SBC Pr
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5159833 VDARC - Volume Data Acq Recon Control Computer, 2nd Edition (Jarrell). See Long Description for Details.. Part #: 5159833 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: VDARC - Volume Data Acq Recon Control Computer, 2nd Edition FRU: Yes Status: Orderable Parts Offered as: Current : 5159833 Harvested : Not Available Exchange : Not Available Comments: Per Jarrell RSPL, 7/17/06, ms11892. ** FE Assist ** Must order software release 06MW29.7, p/n 5193989 and perform a LFC to support this FRU if system is not currently running 06MW29.7 or better.(12/8/06 212001261
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.