Symptom

Collimator calibration drift check has failed Cal will continue with the next step - FPA Check.

Error message

Error: 230023045 The Collimator calibration drift check has failed See the long description of the full error code Thu Aug 26 02:56:40 2004 Error: 230023045 Host: ct01 Process: dailyPrepRx File: DailyPrepUtils.c Line: 3155 The Collimator calibration drift check has failed. Full collimator calibration of all stations must be done. Cal will continue with the next step - FPA Check.

Error: 230017801 ACCS CCal: Sweep Processing failed. Status = -1 See the Long Description for the full error code Tue Nov 30 02:33:04 2004 Error: 230017801 Host: k2_ct2_ls Process: accs File: ccal_prep.c Line: 661 ACCS CCal: Sweep Processing failed. Status = -1

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.