Symptom

Gantry tilts slowly 1 degree and stops - table moves very slowly 5mm and stops. Gantry tilts slowly 1 degree and stops - table moves very slowly 5mm and stops

Error message

Error: 260112014 Touch sensor ? Touch Sensor was PRESSED See Long Description for full error Error: 260112014 Touch sensor ? Touch Sensor was Pressed. Sun Mar 12 15:14:25 2006 Error: 260112014 Host: tgp Process: tgp_subsys File: uif_table_tilt_mgr_gt.c Line: 9013 Sun Mar 12 15:14:24 2006 Host : Table Gantry Processor Ermes # : 260112014 Exception Class : Secondary Severity : Sec/Soft File : uif_table_tilt_mgr_gt.c Line# : 9013 Function : PATIENT ANATOMY POSITIONING : Tilt Drive Scan Type : None/Unknown Scan: 0/0/0 Touch sensor BACK Touch Sensor was PRESSED. Elevation position where touch Sensor was Pressed:0 um. Tilt position where touch sensor was pressed:0 mdeg. Remote Tilt button is DISABLED. Use local gantry controls to tilt the gantry. Or Mon Apr 9 09:29:44 2007 Error: 260112014 Host: tgp Process: tgp_subsys File: uif_table_tilt_mgr.c Line: 9990 Mon Apr 9 09:29:44 2007 Host : Table Gantry Processor Ermes # : 260112014 Exception Class : Secondary Severity : Pri/Soft File : uif_table_tilt_mgr.c Line# : 9990 Function : PATIENT ANATOMY POSITIONING : Elevation Drive Scan Type : None/Unknown Scan: 317/6/0 Touch sensor FORWARD Touch Sensor was PRESSED. Elevation position where touch 

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2277404 Patient Tilt Sensor, Large for Rear Gantry Cover used on HiSpeed QXi, 2.x(Plus), 3.x(Ultra), 4.x(16 Slice), HPower and VCT gantries See Long Description for Part Details 2277404 Patient Tilt Sensor, Large for Gantry REAR Cover Asm. Part #: 2277404 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: PATIENT TILT SENSOR, LARGE FRU: Yes Status: Orderable Comments: FLS - H2 GANTRY. COMMON TO MFG MS26561 DSM 1/7/01 REAR LARGE SWITCH/PART OF 2273016 REAR GANTRY COVER COLLECTION OF DEFECTIVES START7/23/01 MS5277
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.