Pular para o conteúdo principal
GE Healthcare · CTHiSpeed18318465001183342

HiSpeed CTi 6.X with Octane and HSA Gantry

Symptom

Scanner Hardware Stopped Scan with High Voltage errors

Error message

Error: 183184 mA out of 5 % tol Open Long Description for full error Tue Jun 7 03:35:46 2005 Error: 183184 Suite: GEC2 Host: OBC Process: Generator File: gen_tech.c 1.36 Method: No Method Line: 924 Function: XRAY GENERATION : MA Loop Scan: 65001/11/14 Type: Axial Exception Level: Sec/Soft Time: 03:35:45:600 Log Series: 223 mA out of 5 % or 2 mA tol Selected technic: 140 kV, 56 mA Tot. samples = 4 Tot. mA samples above tol. = 0, below tol. = 1 Avg. of mA samples above tol. = 0.0 mA, below tol. = 0.2 mA Avg. of tot. mA = 26.7 mA

Error: 183342 Two consecutive anode mA samples out 10.0 % or 4 mA tolerance. See LD for more details Mon May 29 06:45:36 2006 Error: 183342 Host: OBC Process: Generator File: gen_tech.c 1.34 Line: 626 Function: XRAY GENERATION : MA Loop Scan: 9571/2/3 Type: Axial Exception Level: Pri/Most Severe Time: 06:45:36:370 Log Series: 69 Two consecutive anode mA samples out 10.0 % or 4 mA tolerance. Selected technic: 120 kV, 90 mA Sample1: Anode = 163.6 mA Cathode = 88.9 mA. Sample2: Anode = 165.2 mA Cathode = 89.4 mA. Possible cause: mA monitoring circuit failure.

Applies to

Models
HiSpeed
Product
HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
D3101T / D3102T X-ray Tube Used In CT/i and HiSpeed Advantage Part #: D3102T Part Type: GP TUBES Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: TUBE HSA/CTI FRU: Yes Status: Orderable Comments: FLS - NEW 1994 TUBE CAT. FOR RELOAD REPLACES D1199ZR (RELOADED) .9 X .7FS, 1.2-1.2 FS HISPEED ADVANTAGE DIMS 37 X 27 X 28 (LWH) MAXIRAY EQUIVALENT IS D3109T TUBE FAN= 46-219465P1 AM SOURCE = GEMS-MM (MEXICO) *** FLS Tube Names Systems Compatible Heat Capacity Focal Spot Catalog # Model # Insert # Casing # Tube Unit # HSA(Zeus) HiSpeed CT/I, or HiSpeed Advantage 3.5MHu 0.9x0.7/1.2x1.2 D3102T MX165 46-309300G2 46-309500G2 46-309500G801 Tubes started shipping half machined, half cast casings in early 2005. This was stated to be a tempory change in process while a new vendor was being evaluated.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.