Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2005200152000004164294967295200000415200000424

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Less image space available than normal. See long description for examples. Less image space available than normal, even after removing all exams in Image Works. Example: H3 or GOC1 console should have 20,000 (512 images) available on a clean system with an Octane cpu with two 9 GB disk drives. But with the replacement of the 9GB drives with 18 and 73GB drives, image space may increase up to 130,000 (512 images) depending on software. Example: GOC2 console should have 100,000 (512 images) available on a clean system with an HP-8000 cpu with one 18GB disk and one 73GB disk. Example: GOC3 or 4 console should have 200,000 (512 images) available on a clean system with an HP-8200 cpu with one 36GB disk and two 73GB disks. Data above is approximate, but can be used to determine if there is a problem with disk space.

Error message

Error: 200520015 Updating image space, received a bad status code: See the long description for different status codes Sat Jan 22 18:23:06 2005 Error: 200520015 Host: chr09bc1cti Process: environmentMgr File: ImageSpaceFormMgr.c Line: 117 Updating image space, received a bad status code: 200000416 Reason: Response from SdC timeout you may also see this error with different status codes, such as Fri Jul 28 05:12:48 2006 Error: 200520015 Host: vct1 Process: environmentMgr File: ImageSpaceFormMgr.c Line: 121 Updating image space, received a bad status code: -1 Reason: OS error : -1 --- Unknown error 4294967295 Wed Nov 16 08:23:54 2005 Error: 200520015 Host: gect_qxi Process: environmentMgr File: ImageSpaceFormMgr.c Line: 121 Updating image space, received a bad status code: 200000415 Reason: Invalid data from SdC.

Error: 200000424 Image available space failed. See the long description for the full error Sat Jan 22 18:25:36 2005 Error: 200000424 Host: chr09bc1cti Process: dms File: AvailSpace.cxx Line: 155 Image available space failed. code: 200000416 Or Sat Oct 28 06:55:39 2006 Error: 200000424 Host: ct02 Process: dms File: AvailSpace.cxx Line: 155 Image available space failed. Error

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
308I.2_H3.1M5 for LightSpeed 3.X (Ultra, Plus, QX/I with H3 Console Assembly (2266832-3, 2304732-3) or GOC1-Octane2 Console (2341104, -2) - showprods returns LightSpeed Apps ver 308I.2_H3.1M5 308I.2_H3.1M5 for LightSpeed 3.X (Ultra, Plus, QX/I with H3 Console Assembly (2266832-3, 2304732-3) or GOC1-Octane2 Console (2341104, -2) - showprods returns LightSpeed Apps ver 308I.2_H3.1M5 308I.2_H3.1M5 for LightSpeed 3.X (Ultra, Plus, QX/I with H3 Console Assembly (2266832-3, 2304732-3) or GOC1-Octane2 Console (2341104, -2) - showprods returns LightSpeed Apps ver 308I.2_H3.1M5 Class C Application Software Set CD Part No. Ver. CD Label CD Part No. Ver. CD Label Set Part No. Apps. CD Part No Apps. CD Label showprods/ Service Desktop Intro. Date Comment 2395172 1104I.2_ H4.0M5 CLASSC LightSpeed MDAS 4/8/16/ HiSpeed QXi Class C Software 2395177 2395162, 2395163, 2399867 LightSpeed Plus/ Ultra MDAS Apps Software (Disk 1, Disk 2 and Patch A) 308I.2_H3.1M5 FW47-03 LS2002M5. IRIX Only. Class C/M Common w/ LS 2X, 3X, & 4X.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues