Pular para o conteúdo principal
GE Healthcare · CTLightSpeed244552178517244540178575212002267

LightSpeed 4.X - GRE

Symptom

Scanner Hardware Reset Failed. See LD Scanner Hardware Reset Failed Hardware Failed to reset message appeared on monitor AFTER boot-up WITH Error:178568 Failed to establish socket connection to STC in 10 seconds and errors 244553 and 244554 and power up self test fails A, B, and 9 on STC CPU board The following errors were in the gesyslog Thu Jan 13 04:18:32 2005 Error: 178568 Suite: RP25 Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to STC in 10 seconds Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset. Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC. Also saw failures A, B, and 9 on the power-up test on the STC CPU .

Error message

Error: 244552 ETC controller has failed to reset See the long description for the full error code Wed Aug 24 06:48:26 2005 Error: 244552 Suite: C408 Host: SBC0 Proc: fwmgr File: HwSetUp.m 1.11 Method: hardReset Line: 416 The ETC controller has failed to reset

Error: 178517 Failure writing to socket connected to controller ETC. No route to host See LD for complete error Fri Oct 8 07:49:33 2004 178517 4 ct03 fwmgr NBSSocket.c 690 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failure writing to socket connected to controller ETC. OS error : 113 --- No route to host Error: 178517 Failure writing to socket connected to controller ETC. No route to ho

Error: 244540 Scan Seq Id: 0 Immediate Request cannot be honored since the firmware is in the DOWN state See Long Description for complete error Sun Sep 26 08:36:46 2004 Error: 244540 4 000ctge04 fwmgr FwMgrEventHandler.c 1517 Running diagnostic: Scan Data Path Functional Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Immediate Request cannot be honored since the firmware is in the DOWN state

Error: 178575 Failed to R-HARD the following controllers: ETC. 

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-264368G3 ETC board See long description for details Part #: 46-264368G3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: ENHANCED TABLE CONTROL HELIOS 2 FRU: Yes Status: Orderable Comments: Not Available
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required