Symptom
HP Host computer intermittently locks up or will not boot up to applications HP 8000 or 8200 computer intermittently locks up or will not boot up to applications
Error message
kernel: NVRM: Xid error in messages file. see long description. Extracts of /var/log/messages at the time of console slowdown or crashes. The NVRM Xid errors were suggested to come from the Graphic board nVidia Quatro 980XGL. Sep 14 22:34:37 ls16 cupMonitor: CUP ERROR in cupIpc.c line 590 pid 4257: (null) failed to send on IPC_CHANNEL Sep 14 22:34:41 ls16 pam_rhosts_auth[5800]: allowed to ctuser@kfhm_oc as ctuser Sep 14 22:34:41 ls16 rsh(pam_unix)[5800]: session opened for user ctuser by (uid=0) Sep 14 22:34:41 ls16 kernel: NVRM: Xid: 8, Channel 00000006 Sep 14 22:34:41 ls16 kernel: NVRM: Xid: 13, 0000 02004a00 0000004a 00000404 0069000e 00004000 Sep 14 22:34:41 ls16 kernel: NVRM: Xid: 3, C 00000001 SC 00000003 M 00000300 Data 0000000a Sep 14 22:34:42 ls16 rpc.mountd: authenticated unmount request from darc:32774 for /usr/g (/usr/g) Sep 14 22:34:52 ls16 kernel: e1000: eth0: e1000_watchdog: NIC Link is Up 100 Mbps Full Duplex Sep 14 22:34:55 ls16 kernel: NVRM: Xid: 7, Ch 00000000 M 00000104 D 00000000 intr 00010000 Sep 14 22:35:11 ls16 afViewer: CUP ERROR in cupMakeCore.c line 198 pid 5717: generating traceback file traceback.afViewer.200614Sep.223511
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2368697-3 HEWLETT-PACKARD HP8000 Computer used in Global Recon Engine (GRE) CONSOLE 2368697-3 GRE CONSOLE HP8000 2368697-3 HP8000 COMPUTER USED IN GRE CONSOLE Part #: 2368697-3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: LINUX DDR WORKST'N PS2 MOUSE--100 PC FRU: Yes Status: Orderable Comments: 3/4/05 (212001261 Shawn Fox) HOST COMPUTER (HP). THISIS THE GOC2 HP 8000 COMPUTER. ***EXTREMELY IMPORTANT: NEW SOFTWARE, S/W LOAD PROCEDURE AND S/W OPTION KEYS ARE REQUIRED IF THE HP8000 COMPUTER IS REPLACED. Please contact the OLC for assistance with new Option Keys
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.