Pular para o conteúdo principal
GE Healthcare · CTLightSpeed183124

LightSpeed 4.X - GRE

Symptom

Scanner would not scan on large or small filament

Error message

Error: 183124 FIL/MA fault detected: Inoperative filament inverter. See LD for more details Mon Aug 7 08:55:55 2006 Error: 183124 Host: OBC Process: Generator File: gen_abort.c 1.31 Line: 385 Function: XRAY GENERATION : Filament Selection Scan: 849/2/6 Type: Helical Exception Level: Pri/Most Severe Time: 08:55:55:030 Log Series: 632 FIL/MA fault detected: Inoperative filament inverter. Desired focal spot = 0. (0 = large, 1 = small) Fault latch address = ffdf4d hex, Bit = D7 Mon Aug 7 09:44:27 2006 Error: 183124 Host: OBC Process: Generator File: gen_abort.c 1.31 Line: 385 Function: XRAY GENERATION : Filament Selection Scan: 850/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 09:44:27:210 Log Series: 638 FIL/MA fault detected: Inoperative filament inverter. Desired focal spot = 1. (0 = large, 1 = small) Fault latch address = ffdf4d hex, Bit = D7

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-170021P104 : 10 Amp 700vdc Fuse - used in HEMRC and Axial Dynamic Brake and others See Long Description for more info 46-170021P104 10 Amp 700vdc F1 Fuse -used in HEMRC and Axial Dynamic Brake and others many pictures below Part #: 46-170021P104 Part Type: GP CONSUMABLES Part Weight: 11.81 LBR Repairable: No Returnable: No Hazardous Material: N Description: FUSE 10 AMPS 700 VOLTS FRU: Yes Status: Orderable Comments: FLS - FUSE 10 AMPS 700 VOLTS ULTRA FAST. USED ON FAST RHAPSODE 7/14/97 GWENN 08992 Fuse for Axial Brake Module Fuse for HEMRC Rotor controler (HSA-1.x) Fuse for HEMRC Rotor Control (2.x-4.x)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.