GE Healthcare · CTBrightSpeed26010009350164
BrightSpeed HPower or Delight HPower
Symptom
Scan fails aborts with PDU HVDC Buss Fault Status Scan fails aborts with PDU HVDC Buss Fault Status.
Error message
Error: 260100093 PDU HVDC Buss Fault Status. The HVDC did not charge correctly in the alotted time Mon Sep 5 13:55:52 2005 Error: 260100093 Host: tgp Process: tgp_subsys File: scIsrService.c Line: 246 Running diagnostic: kV and mA Test: Xray Functional Test Mon Sep 5 13:55:51 2005 Host : Table Gantry Processor Ermes # : 260100093 Exception Class : Abort Severity : Primary File : scIsrService.c Line# : 246 Function : SYSTEM MONITORING : HVDC Monitor Scan Type : Static Scan: 50164/1/1 PDU HVDC Buss Fault Status. The HVDC did not charge correctly in the alotted time. Threshold 1: > 200V in 200ms Threshold 2: > 500V in 1.2s. Possible causes: HVDC Cable Short Circuit, SlipRing, or Jedi Inverter.
Applies to
- Models
- BrightSpeed
- Product
- BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2113764-3 : Capacitors 6uF 370VAC (not 6NF) See long description for details 2113764-3 : Capacitors 6uF 370VAC (not 6NF) Part #: 2113764-3 Part Type: GP CONSUMABLES Part Weight: 1.10 LBR Repairable: No Returnable: No Hazardous Material: N Description: FILTER CAPS FRU: Yes Status: Orderable Parts Offered as: Current: 2113764-3 Harvested: Not Available Exchange: Not Available Comments: FLS - THERE IS A KNOWN PROBLEM WITH THE DESIGN AND REPLACING CAPS WILL DO IT AGAIN,FMI IS BEING WORKED ON,CONTACT CT SUPPORT BEFORE ORDERING PER JOHN SHUBERT 9-24-96 FILTER CAP C1,C2,C3,C7,C8,C9
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required