LightSpeed 5.X (60kW)

Symptom

"Scan hardware Stopped Scan" with Axial Errors. "Scan hardware Stopped Scan" with Axial Errors

Error message

Error: 230023053 Failed on Gantry Firmware Mon Feb 4 07:19:52 2008 Error: 230023053 Host: aafgect1 Process: dailyPrep Rx File: DailyPrepUtils.c Line: 5433 Failed on Gantry Firmware. Cannot continue.

Error: 260107404 The Axial AC Drive reported fault code:F4- Undervoltage Fault. The Drive internal DC bus voltage dropped below 325V. Possible low voltage condition on 120VAC in gantry or power interruption. See long description for full error Thu Feb 22 18:00:09 2007 Error: 260107404 Host: tgp The Axial AC Drive reported fault code:F4- Undervoltage Fault. The Drive internal DC bus voltage dropped below 325V. Possible low voltage condition on 120VAC in gantry or power interruption.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.