Pular para o conteúdo principal
GE Healthcare · CTLightSpeed1831212257550013

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Could not scan on small filament

Error message

Error: 183121 FIL/MA fault detected: Open filament circuit. See the long descripton for the full error code Tue Oct 11 13:12:27 2005 Error: 183121 Host: OBC Process: Generator Function: XRAY GENERATION : Filament Selection Scan: 22575/2/1 Type: Helical Exception Level: Pri/Most Severe FIL/MA fault detected: Open filament circuit. Desired focal spot = 0. (0 = large, 1 = small) <== Large Filament is open OR Fri Sep 22 07:03:11 2006 Error: 183121 Host: OBC Process: Generator File: gen_abort.c 1.31 Line: 341 Function: XRAY GENERATION : Filament Selection Scan: 50013/10/40 Type: Axial Exception Level: Pri/Most Severe Time: 07:03:11:045 Log Series: 393 FIL/MA fault detected: Open filament circuit. Desired focal spot = 1. (0 = large, 1 = small)<-------- Small

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-136370P23 K78 FILAMENT RELAY See long description for details Part #: 46-136370P23 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: 024VDC 2P2T 028VDC 10AMP FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues