Pular para o conteúdo principal
GE Healthcare · MR30000042248349

New Patient Screen Not Available Due to Corrupted prescan_config.cfg File.

Symptom

New Patient Screen Not Available Only thing visible on IDLE screen are the 6 Desktop Icons Upper Left Corner Only thing visible on IDLE screen are the 6 Desktop Icons Upper Left Corner.

Error message

3000004 An exception has occurred in MxJVM Mon Mar 6 05:57:20 2006 Error: 3000004 Host: GH-GEMR Process: LxMxConsumer File: MxJVM Line: 0 An exception has occurred in MxJVM.^C Exception: Mon 03/06/2006 05:57:20.672 AM EST::::Stack Trace Follows: Exception: jav

2248349 The file prescan_config.cfg contains a missing or bad value for hos_exists. Apr 20 Fri 06:53:39 2248349 NSP SCP:scp_set_shim_config 5346 The file prescan_config.cfg contains a missing or bad value for hos_exists.

Applies to

Product
Open Speed with EXCITE For HFO4 and HFO5 only
Software
OPENMR5-HFO5 Open Speed with Excite Product
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues