Pular para o conteúdo principal
GE Healthcare · CTLightSpeed

LightSpeed 4.X - Octane

Symptom

System will not boot to applications WITH Media error: Unrecovered data block read error.

Error message

dksc0d2vol: 6 Device spin up failed, unable to use device -- corrective action necessary See long description for details Open unix shell and login as root cd /usr/adm more SYSLOG Use space bar to page through log and look for disk drive errors Apr 23 14:00:59 6A:bdicnd ls16 unix: dksc0d2s0: 6 retries exhausted Apr 23 14:00:59 1A:bdicnd ls16 unix: ALERT: I/O error in filesystem ("/usr/g/sdc image pool2") meta-data dev 0xee block 0x800040 Apr 23 14:50:27 6A:bdicnd ls16 unix: dksc0d2vol: 6 Device not ready 6 : Not ready to perform command (asc=0x4, asq=0x0) 6 Apr 23 14:50:27 6A:bdicnd ls16 unix: dksc0d2vol: 6 Device spin up failed, unable to use device -- corrective action necessary

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2169940-81 Octane Computer 73 GB Disk Drive - replaces 2169940-78 (18GB) See long description for aditional information 2169940-81 Octane 73 GB Disk Drive, replaces 18 GB Disk 2169940-78 Part #: 2169940-81 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: CT OCTANE 2 73 GB DISK DRIVE FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.