Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 182226 Warning: Cradle did NOT reach Scan Position. see Long Description for complete error Fri Sep 1 15:04:38 2006 Error: 182226 Host: ETC Process: Cradle File: cmcShutDown.c 1.54 2/5/96 Line: 184 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 1145/2/5 Type: Axial Exception Level: Pri/Soft Time: 15:04:38:470 Log Series: 449 Warning: Cradle did NOT reach Scan Position. Start Position 33542 (.01 mm) Scan/End Position 32044 (.01 mm) Distance (Start-Scan) 1498 (.01 mm) Current Position 25299 (.01 mm) Tolerance 25 (.01 mm)

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2269169 CRADLE DRIVE ASM FOR CATHAY TABLE. See Long Description for details Part #: 2269169 Part Type: GP REPAIRABLES Part Weight: 14.23 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: CRADLE DRIVE ASM FOR CATHAY TABLE FRU: Yes Status: Orderable Comments: FLS - CRADLE DRIVE ASM FOR CATHAY TABLE. P/N 2269169 REPLACES P/N 46-296235G2 PER PCN 220590USE FIRST EXISTING STOCK. MS20544 ON 8/23/02.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.