GE Healthcare · CT23000011423000071323000006750000245190
Lightspeed 1.X - Octane
Symptom
Scanner Hardware Stopped Scan with DIP and view length errors. Scanner Hardware Stopped Scan with DIP and view length errors
Error message
Error: 230000114 DIP scan relay test: Initial clearing of DIP interrupt status register failed See the long description for the full error code Mon Aug 1 00:50:35 2005 Error: 230000114 Host: qxi1_sbc Process: dataacq File: dip_if.cxx Line: 1850 Function: Data Acquisition : SBC Processing Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 A DIP error was detected. DIP scan relay test: Initial clearing of DIP interrupt status register failed. Failure code: 230000713 Test code: 00080001h Address: 00000003h Expected value: 00000000h Actual value: 00000008h Error: 230000067 View Length error See LD for more details Tue Aug 22 12:35:04 2006 Error: 230000067 Host: ct01_sbc Process: dataacq File: dip_if.cxx Line: 416 Function: Data Acquisition : Scan Data Acquisition Scan Type: Axial Scan: 50000/1/4 Scan Seq Id: 4000 Unexpected event occurred during scan data transfer. Event bit: 5 Count: 1 Description: View length error Scan aborted. Error: 245190 System aborted Axial scanning. No views, either offset or data, were collected/received. Data views expected: 992 See the Long Description for the full error code. Thu Jan 13 06:58:23 2005 Error: 245190 Host: ct02 Process: scanmgr File: scan_c
Applies to
- Product
- Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2167012 RF Receiver - for old style ETC slip ring used on 1.x, YMS & Ct/e If you have the newer HSDCD slip ring this part IS NOT FOR YOU - HSDCD uses 2245896 rf receiver See Long Description for picture If you have the newer HSDCD slip ring this part IS NOT FOR YOU - HSDCD uses 2245896 rf receiver Part #: 2167012 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.99 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: P9237HM RF SLIPRING RECEIVER FRU: Yes Status: Orderable Item 1-12 is the RF Receiver, p/n 2167012. Direction 2188542, Rev.40, Section 2-2-1-3.
- Area
- CT
Full cause & step-by-step resolution