Symptom

Scanning fails with recon errors

Error message

Error: 290010004 Checksum mismatch for view 1241. Expected 204632862 : Got See Long Description Tue Dec 6 11:07:48 2005 Error: 290010004 Host: darc Process: data_restore File: mdas_intf_object.cxx Line: 372 Checksum mismatch for view 1241. Expected 204632862 : Got 204633118

Error: 290009001The scanfile does not have enough views to make requested images. See LD Error: 290009001The scanfile does not have enough views to make requested. Tue Dec 6 11:07:48 2005 Error: 290009001 Host: darc Process: recon_control File: ImageSetRequest.cxx Line: 2684 The scanfile does not have enough views to make requested images. exam/series/scan = 60/2/1 expected number of images = 12 created number of images = 0 max number of images = 12 scan completed = 1

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5144058-500 Upgrade kit - GOC3 to GOC4. DARC2 is not included. Order proper DARC2 separately - REFER to SN11063012 See Long Description for Details. 5144058-500 Upgrade Kit GOC3 to GOC4 - without DARC2 Part #: 5144058-500 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Service Kit - SDDA and GOC3 16 MDAS 16,8,4 GDAS DARC Replacement kit, JARRELL FRU: Yes Status: Orderable Comments: ** FE ASSIST ** This part number no longer contains a DARC2 it must be ordered in addition per Your system type. REFER to SN11063012a(July 26 2011 212001261)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.