GE Healthcare · CT
Lightspeed 1.X - Octane
Symptom
System will not boot to or stay in applications. System will not boot or stay in applications
Applies to
- Product
- Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2169940-72UU (replaces 2169940-72) : Octane Computer 9 GB System Disk See Long Description for details Note: There are compatability problems with some consoles and disk drive firmware. Please see service note SN07112612a for details. Part #: 2169940-72UU Part Type: GP USED Part Weight: 2.43 LBR Repairable: No Returnable: No Hazardous Material: N Description: 9GB 10K RPM DISK WITH SLED FRU: Not Available Status: Orderable Comments: NOTE: MAY BE INACCURATE FOR H2 AND H3 CONSOLES. ** FE ASSIST ** 2169940-12 is no longer available can use 2169940-72 9GB or a 2169940-78 18GB Hard drives for a CTi or lightspeed QXI Octane systems as Image or OS disk. Will need to use updated LFC per 2282382-100 rev 2 or 2284291-100 rev 8. The 2169940-78 has 5000 images files where normal is 20000 if used as a CTi Image Disk. Refer to Service Note 07112612b. (212001261 8/11/09) ***OBSOLETE PART Case # 9746502*** This part has been identified as technically obsolete and no short term solution has been identified. FE must identify and order a higher or lower level spare as required. Contact your support engineering, On-line Engineering or Parts ID team should you require support. 212012173 7-24-236 Part #: 2169940-72 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 1.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: 9GB Octane 10K RPM DISK WITH SLED FRU: Yes Status: Orderable Parts Offered as: Current: 2169940-72 Used: Not Available Exchange: Not Available Comments: OCTANE 2 9GB DISK DRIVE USED ON HELIOS 2 AND 3 AND LX 9/25/03 CHECK WITH FIELD TO SEE IF THEY CAN USE THE 18GB PART# 2169940-78 IF NONE OF THE 9GB ARE AVAILIABLE
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required