Symptom

Scan abort during alpha calibration. Scan abort during alpha calibration

Error message

Error: 260007936 DAS Converter Board(s) encountered an error during alpha calibration. See LD for complete error which shows failing board number Error: 260007936 Host: orp Process: orp_subsys File: \ Host : Das Ermes # : 260007936 Exception Class : Soft Severity : Pri/Soft File : brdflt.c Line# : 82 Function : No System Function Reported Scan Type : None/Unknown Scan: DAS Converter Board(s) encountered an error during alpha calibration. See Following Error Message for Converter Card List Board :6 ASIC/Channel :0x0137 Stage :1 Board Channel :19

Error: 260007402 Converter Board(s) Asserted a Board FAULT. See Long Description for complete error Tue Dec 28 09:41:31 2004 Error: 260007402 Host: orp Process: orp_subsys File: cnvmgr.c Line: 605 Tue Dec 28 09:41:30 2004 Host : Das Ermes # : 260007402 Exception Class : Abort Severity : Secondary File : cnvmgr.c Line# : 605 Function : No System Function Reported Scan Type : Static Scan: 50444/1/1 Converter Board(s) Asserted a Board FAULT.

Applies to

Models
BrightSpeed
Product
BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2318808-3 150 PIN - 5 connector INTERPOSER Replaces part 2318808-2 used in MDAS-16 and GDAS-16 (Quantity = 5) 2318808-3 150 PIN - 5 connector INTERPOSER Replaces part 2318808-2 used in MDAS-16 and GDAS-16 systems. Part #: 2318808-3 Part Type: GP CONSUMABLES Part Weight: 1.09 LBR Repairable: No Returnable: No Hazardous Material: N Description: (5) CONNECTORS, 150 PIN INTERPOSER FRU: Yes Status: Orderable Comments: FLS - CONNECTORS (5). ADDED PER CT PCN 220814. REPLACES P/N 2318808-2 SINCE THIS -2 PART CHANGE FROM FRU YES TO NO. MS20544 5/20/03 PCN NOTES: 2318808-3 IS A TUBE THAT CONTAINS 5 PIECES. FOR QUANTITY 1 ORDERED FE WILL RECEIVE 5 INTERPOSERS. VENDOR INFO MILL-MAX MFG. CORP S596-00-000-00-955000 VC 462905 8536.69.4040 HTS CODE
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.