GE Healthcare · CTBrightSpeed26000777426000740250444260007770260007939260101023
BrightSpeed 4/8/16 slice HPower Scanner Elite-16, Edge-8, Excel-4
Symptom
Multiple rings and bands (Blown Image) image artifacts with DAS errors See Long Description for Image Multiple rings and bands (Blown Image) with DAS errors GDAS 16
Error message
Error: 260007774 List of Converter Cards From the Previous Error Message Card List: xx xx See Long Description for full error Tue Dec 30 09:25:17 2008 Error: 260007774 Host: orp Process: orp_subsys File: CnvBoardInterface.cc Line: 852 Tue Dec 30 09:25:17 2008 Host : Das Ermes # : 260007774 Exception Class : Secondary Severity : Sec/Soft File : CnvBoardInterface.cc Line# : 852 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 List of Converter Cards From the Previous Error Message Card List: 8,11,14-15,19,21,25-27,29-36,38-58,60-63 Note on this error. Sometimes the actual bad board will be the one that is not listed. Error: 260007402 Converter Board(s) Asserted a Board FAULT. See Long Description for complete error Tue Dec 28 09:41:31 2004 Error: 260007402 Host: orp Process: orp_subsys File: cnvmgr.c Line: 605 Tue Dec 28 09:41:30 2004 Host : Das Ermes # : 260007402 Exception Class : Abort Severity : Secondary File : cnvmgr.c Line# : 605 Function : No System Function Reported Scan Type : Static Scan: 50444/1/1 Converter Board(s) Asserted a Board FAULT. Error: 260007770 : Converter Board Broadcast Communication Error._ See long description for full error Er
Applies to
- Models
- BrightSpeed
- Product
- BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2361853-2 GDAS DCB 16-SLICE replaces 2361853 See LD for more Info 2361853-2 GDAS DCB 16-SLICE replaces 2361853 Part #: 2361853-2 Part Type: GP CONSUMABLES Part Weight: -59999.40 LBR Repairable: No Returnable: No Hazardous Material: N Description: GDAS DCB FRU: Yes Status: Orderable FLS - GDAS DCB (16-SLICE) USED ON: LIGHTSPEED PRO 16 GDAS ADDED PER GDAS RSPL MS13452 This board has been replaced by 2361853-2, FE NOTE: Site MUST have M4FMI completed before this board will function properly, ms13452
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required