GE Healthcare · CTLightSpeed230017822850000830000
LightSpeed 5.X (80/100kW)
Symptom
Collimator calibration drift check has failed Cal will continue with the next step - FPA Check.
Error message
Error: 230017822 ACCS CCal: Collimator calibration failed. See long description for full error Thu Jun 5 07:21:16 2008 Error: 230017822 Host: ct01 Process: accs File: collSweepProcessing.c Line: 1417 ACCS CCal: Collimator calibration failed. The measured aperture 3.850000 exceeds the upper limit of 3.830000 for this mode. The database will not be updated
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution