Pular para o conteúdo principal
GE Healthcare · CTLightSpeed23001707850342111152230710858023001708650055

LightSpeed 5.X (60kW)

Symptom

DAILY IQ CHECK THRESHOLD FAILURE

Error message

Error: 230017078 Aircal failed to process scan See Long Description for complete error Tue Mar 22 20:12:04 2005 Error: 230017078 Host: ct01 Process: accs File: accs_events.cxx Line: 463 Aircal failed to process scan - Exam = 50342, Series = 1, Scan = 1, Seconds = 1111522307 Microseconds = 108580 View scanfileto check data and header information

Error: 230017086 DAILY IQ CHECK THRESHOLD FAILURE See Long Description for the full error Error: 230017086 Host: mcop_sbc Process: accs File: aircal_processing.c Line: 797 DAILY IQ CHECK THRESHOLD FAILURE - Scan 50055/8/3 Drift check fail Kv - KV_80, Sfov - SFOV_LARGE, Colimation - SLICE_COLLIMATION_4x125, Focal spot -FOCAL_SPOT_SMALL, Das Gain - Value:30, Row - 0, Channel - 467, Baseline mean - 0.153948, Current mean - 0.154175, Smoothed ratio - -0.007852, Lower limit - -0.005000, Upper limit - 0.005000 .

Error: 230015028 Failed to construct a file path for vector type 2 See long description for complete error Fri Feb 2 06:16:14 2007 Error: 230015028 Host: ls16 Process: accs File: cal db api.c Line: 1893 Function: CAL Processing : SBC Processing Failed to construct a file path for vector type 2 in directory /usr/g/cal db/Value:48/large sf

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2404393 GDAS-16 CONVERTOR CARD, replaces 2364354. See Long Description for details. 2404393 GDAS-16 CONVERTER CARD, replaces 2364354 Part #: 2404393 Part Type: GP CONSUMABLES Part Weight: 2.20 LBR Repairable: No Returnable: No Hazardous Material: N Description: 16 SLICE GDAS CONVERTER CARD FRU, A CONVERTER CARD IN AN ESD BAG IN A BOX WITH A PART NUMBER LABEL FRU: Yes Status: Orderable Comments: FLS - 16 SLICE GDAS CONVERTER CARD USED ON: LIGHTSPEED PRO 16 FRU GDAS ADDED PER GDAS RSPL MS13452 040315 RISK: SITE DOWN FREQUENCY: HIGH NOTE: THERE ARE 96 OF THESE CARDS PER SYSTEM, BUT SHOULD NOT NEED TO REPLACE ALL AT ONETIME, MS13452 THIS PART CONTAINS ONE GDAS CONVERTER CARD 2364354 WITH FRU PACKAGING, SUPPLIED BY GEMS IT. MS13452 PDR 8471.80.9000 IM05 UPDATE AWAITING MM (212001261 4/1/06) For 4/8 slice G Das which uses a thick connector must order 2364354-2
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.