GE Healthcare · CTLightSpeed230011000230019002
LightSpeed 2.X (Plus, QX/i) - Octane
Symptom
System will not stay up in applications and shutsdown. System will not stay up in applications and shutsdown
Error message
Error: 230011000 Scan data disk handling not initialized. A successful initialization of the Scan Disk I/O subsystem must occur prior to writing or accessing scan data See LD for the full error Tue Aug 23 08:03:24 2005 Error: 230011000 Host: ice Process: sdiomgr File: sdiomgr.cxx Line: 2315 initRawDataDisk failed Scan data disk handling not initialized. A successful initialization of the Scan Disk I/O subsystem must occur prior to writing or accessing scan data Error: 230019002 The scan disk I/O layer failed to consume a response from the scan disk server. See LD for complete error Tue Nov 2 03:19:30 2004 Error: 230019002 Host: sgredct1 Process:scanfilemgr File: sdio proxy.c Line:174 The scan disk I/O layer failed to consume a response from the scan disk server. function:consume sequence:6708
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2370825 RIP / Motorola Board, 2400 Series Card - replaces all previous part numbers - Oct 14 2004 See long description and attachment 2370825 RIP / Motorola Board, 2400 Series Card, replaces all previous part numbers - 10/14/2004 Part #: 2370825 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 9.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: PPC MVME 2400 ASSY, HAWK3 U3D 256MB FRU: Yes Status: Orderable Comments: *** OBSOLETE PART *** This part has been identified as obsolete and no short-term solution has been identified. FE must identify and order a higher or lower level spare part as required. Contact your support engineer, on-line engineer or Parts ID team should you require support. 212012173 10-2-06 2/3/05 (212001261) New RIP-Bd 2370825 only works with LS2002M5 on LS 2X/3X & HS_QXi. LS 4X needs to run LS2002M5.1. To confirm what software version is running in the console, open a shell and type in check_config Even if the system is running LS2002M5, a LFC will probably be required. To determine if a LFC is required, open a shell and type in: showprods (the pipe symbol) grep Helios_MVME enter] if the prompt returns (Helios_MVME24XX_VxWorks_Product) where XX = any number, then no LFC, run pflash. If the prompt returns something else, LFC is required. GOC2/Linux-Pegasus are compatible with RIP-Bd P/N 2370825, (pflash) will need to be run. The following s/w versions support RIP Board 2370825: HS QX/I (IRIX) = 2401246 (101I.2_HSQXIM5), LS 2X (IRIX) 2395178 (2.4.2_H2.4M5); LS 3X (IRIX) = 2395177 (308I.2_H3.1M5), LS 4X (IRIX) = 5116050 (405I.3_H4.0M5.1), HS QX/I (Linux) = 2401245 (103L.4_HSQXIM4), LS 3X (Linux) = 2400882 (308L.5_H3.3M4), LS 4X (Linux) = 5111370 (405L.8_H4.2M4). Can order software via Coakly. Alejandro Cortez
- Area
- CT
Full cause & step-by-step resolution