GE Healthcare · CTLightSpeedVCT230007011

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

A pink Attention box pops up during Fast Cal saying there is a problem with the DAS/Detector. Scanning will NOT be disabled. Image Quality will be compromised, call GE SERVICE. See long description for pop up message picture. A pink Attention box pops up during Fast Cal saying there is a problem with the DAS/Detector. Scanning will NOT be disabled. Image Quality will be compromised, call GE SERVICE. After clicking OK Fast Cal completes. ATTENTION: The system detected a problem with the DAS/Detector sub-system. Image Quality will be compromised. Scanning will NOT be disabled, but it is critical that the user call GE SERVICE to rectify the problem. Press 'OK" to proceed at the risk of Image Artifacts

Error message

Error: 230007011 Density Check - Module (X) exceeds number of shorts allowed (X) = modual number See LD for complete error Mon Oct 31 07:03:37 2005 Error: 230007011 Host: cpcvct2p Process: badchanmgr File: DensityMap.cxx Line: 63 Density Check - Module 6 exceeds number of shorts allowed (151 > 64). Sat Aug 15 14:04:26 2009 Error: 230007011 Host: vct Process: badchanmgr File: DensityMap.cxx Line: 63 Density Check - Module 24 exceeds number of shorts allowed (320 10).

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5134959 FRDM Module L-FINE 32 replaces 5123019 Used on VCT 32 Detector See Long Description for Details. 5134959 FRDM Module L-FINE 32 replaces 5123019-Used on VCT 32 Detector Part #: 5134959 Part Type: GP CONSUMABLES Part Weight: -98999.01 LBR Repairable: No Returnable: No Hazardous Material: N Description: L-FINE 32 Detector Module Kit FRU: Yes Status: Orderable Part #: 5123019 Part Type: MST DEFAULT LOAD Part Weight: 0.00 LBR Repairable: Not Available Returnable: Not Available Hazardous Material: N Description: L-FINE 32 DIGITAL MODULE ASSY MP3 FRU: Yes Status: Non Orderable
Area
CT