GE Healthcare · CTHiSpeed18221360046
HiSpeed CT/i 4.X with Indigo and HSA Gantry
Symptom
Grinding noise coming from the table while driving cradle in or out. Table may move slower than normal. Customer may hear a grinding noise coming from the table while driving cradle in or out. Table may move slower than normal.
Error message
Error: 182213 Velocity tracking error See Long Description for the full error code Tue Mar 15 15:28:05 2005 Error: 182213 Host: ETC Process: Cradle File: CmdMotion.c 1.33 7/23/96 Line: 92 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 15:28:05:770 Log Series: 37 HW error: Velocity tracking error. Expected Velocity 500 (.001 mm/sec) Actual Velocity 9382 (.01 mm/sec) Current Position 60046 (.01 mm) Velocity Tolerance 250 (.001 mm/sec) Cradle Dir: 1 (0 = Out 1 = In [Gantry]).
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2269169 CRADLE DRIVE ASM FOR CATHAY TABLE. See Long Description for details Part #: 2269169 Part Type: GP REPAIRABLES Part Weight: 14.23 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: CRADLE DRIVE ASM FOR CATHAY TABLE FRU: Yes Status: Orderable Comments: FLS - CRADLE DRIVE ASM FOR CATHAY TABLE. P/N 2269169 REPLACES P/N 46-296235G2 PER PCN 220590USE FIRST EXISTING STOCK. MS20544 ON 8/23/02.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required