LightSpeed 5.X (80/100kW)

Symptom

Scanning aborted with Buffer overrun detected during scan data transfer

Error message

Error: 230000065 Buffer overrun detected during scan data transfer. See long description for full error Error: 230000065 Buffer overrun detected during scan data transfer. This is generally not a hardware issue and will not be remedied by replacing or otherwise servicing the gantry, fiber-optic cable or DIP card. Interrupt bit: 6 Violation count: 1 Scan aborted.

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-183000P164 10 OZ. SQUEEZE BOTTLE of ALCOHOL 99% 46-183000P164 10 OZ. SQUEEZE BOTTLE of ALCOHOL 99% 46-183000P164 10 OZ. SQUEEZE BOTTLE of ALCOHOL 99%
Area
CT

Full cause & step-by-step resolution