Pular para o conteúdo principal
GE Healthcare · CT26000309750002

Lightspeed 1.X - Octane

Symptom

Fastcals failed to complete. Fastcals failed to complete

Error message

Error: 260003097 Aperture Sweep failed to complete See the long descripiton for the full error code Tue Aug 23 06:24:42 2005 Error: 260003097 Host: HSC Process: CCB Firmware File: aperture_sweep.c Line: 558 Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 50002/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 06:24:41:370 Log Series: 47 Aperture Sweep failed to complete. Steps Requeested: 37 Steps Performed : 35

Applies to

Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues