Pular para o conteúdo principal
GE Healthcare · CTLightSpeed

LightSpeed 4.X - GRE

Symptom

Monitor has a line or other artifact that is not part of the image. If the image is moved or resized the artifact does not move with it. See Short description Monitor has a line or other artifact that is not part of the image. If the image is moved or resized the artifact does not move with it.

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5169069 19 INCHES COLOR MONITOR LCD 1990SXi. See long description for details Part #: 5169069 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 2.43 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: OEM 19 INCHES COLOR MONITOR LCD 1990SXi - SILVER FRU: Yes Status: Orderable Comments: Check graphics board output connector: if 13W3M you need to order 13W3M/HD15 adaptor: 2261178. Per eNPI Project 015197, 212008898, 11/10/06. This monitor should not be ordered for use on 11.X, 10.X, 9.X, or 8.X Signa MR systems until further notice. Required monitor gamma table is not yet available for these systems. Instead order 5117545. (212002950 - 4/25/2007)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues