Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT23001900224528350003

LightSpeed 5.X (60kW)

Symptom

"Scan hardware Stopped Scan" with scan data errors. "Scan hardware Stopped Scan" with scan data errors

Error message

Error: 230019002 The scan disk I/O layer failed to consume a response from the scan disk server. See LD for complete error Tue Nov 2 03:19:30 2004 Error: 230019002 Host: sgredct1 Process:scanfilemgr File: sdio proxy.c Line:174 The scan disk I/O layer failed to consume a response from the scan disk server. function:consume sequence:6708

Error: 245283 Timeout (6000 ms) waiting on scan disk rezero response from dataacq See long description for full error Mon Nov 26 07:29:05 2007 Error: 245283 Host: svmcct16 Process: scanmgr File: scanmgr events.cxx Line: 1151 Function: System Control : Scan Control Scan Type: Not scanning Scan: 50003/2/0 Scan Seq Id: 0 Timeout (6000 ms) waiting on scan disk rezero response from dataacq.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2368697-3 HEWLETT-PACKARD HP8000 Computer used in Global Recon Engine (GRE) CONSOLE 2368697-3 GRE CONSOLE HP8000 2368697-3 HP8000 COMPUTER USED IN GRE CONSOLE Part #: 2368697-3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: LINUX DDR WORKST'N PS2 MOUSE--100 PC FRU: Yes Status: Orderable Comments: 3/4/05 (212001261 Shawn Fox) HOST COMPUTER (HP). THISIS THE GOC2 HP 8000 COMPUTER. ***EXTREMELY IMPORTANT: NEW SOFTWARE, S/W LOAD PROCEDURE AND S/W OPTION KEYS ARE REQUIRED IF THE HP8000 COMPUTER IS REPLACED. Please contact the OLC for assistance with new Option Keys
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues