GE Healthcare · CTBrightSpeed26011818550103260118120
BrightSpeed 4/8/16 slice HPower Scanner Elite-16, Edge-8, Excel-4
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260118185 (50-0204) Filament permanent open circuit See Long Description for complete error. Host: orp Process: orp_subsys File: gen_exec.c Line: 2376 Wed Mar 21 08:25:44 2007 Host : Jedi Ermes # : 260118185 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2376 Function : No System Function Reported Scan Type : Axial Scan: 50103/8/6 (50-0204)Filament permanent open circuit. Issued by the heater board when an open has been detected and 3 restarts have been tried without success within 100 ms. ^C Z6,Z7= 97h,0h Error: 260118120 (20-0255) Filament Open circuit Ermes: 260118120 (20-0255) Filament Open circuit Wed Mar 21 08:25:45 2007 Error: 260118120 Host: orp Process: orp_subsys File: gen_exec.c Line: 2376 Wed Mar 21 08:25:44 2007 Host : Jedi Ermes # : 260118120 Exception Class : Secondary Severity : Pri/Soft File : gen_exec.c Line# : 2376 Function : No System Function Reported Scan Type : Axial Scan: 50103/8/6 (20-0255)Filament open circuit(inverter1)(<3 times). ^C Z6,Z7= 97h,0h
Applies to
- Models
- BrightSpeed
- Product
- BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2361419 Programmed Heater Board (eqvt 2409049) inside Jedi Auxiliary Module 2324118 See LD for Part Details - please read the note at the bottom. Part number for Heater (Filament) Board below: Part #: 2361419 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: PROGRAMMED HEATER 6P04 FRU: Yes Status: Orderable Part #: 2409049 Part Type: GP CONSUMABLES Part Weight: 0.00 KGM Repairable: No Returnable: No Hazardous Material: N Description: FRU-PROGRAMMED HEATER V6.04 FRU: Yes Status: Orderable Parts Offered as: Current: 2409049 Harvested: Not Available Exchange: Not Available Comments: added per to ECR2088514 by sso#305010201 in Jan 15, 2010 Note: * Order 2409049 to replace 2361419.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.