GE Healthcare · CTLightSpeedVCT260007391
LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry
Symptom
"Scan hardware Stopped Scan" with DAS power errors. "Scan hardware Stopped Scan" with DAS power errors
Error message
Error: 260007391The Data Acquisition System Has Detected An Error In The +5V DCB/DAS Digital DAS Power Supply. Wed Apr 5 10:58:29 2006 Error: 260007391 Host: orp Process: orp_subsys File: pwrtst.c Line: 90 Wed Apr 5 10:58:28 2006 Host : Das Ermes # : 260007391 Exception Class : Abort Severity : Secondary File : pwrtst.c Line# : 90 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 The Data Acquisition System Has Detected An Error In The +5V DCB/DAS Digital DAS Power Supply.^B Measured: 0 mV, Expected range: 4750 to 5250 mV
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.