GE Healthcare · CTLightSpeed1831846500118318396082210922
LightSpeed 4.X - GRE
Symptom
CTx00088 Proactive svc call by PDS See long description for details CTx00088 Proactive svc call by PDS This Issue was detected by the Digital Services (DS) rule engine. This Issue is only applicable when detected by the Rules Engine and should not be reused in Knowlege objects for normal reactive service
Error message
Error: 183184 mA out of 5 % tol Open Long Description for full error Tue Jun 7 03:35:46 2005 Error: 183184 Suite: GEC2 Host: OBC Process: Generator File: gen_tech.c 1.36 Method: No Method Line: 924 Function: XRAY GENERATION : MA Loop Scan: 65001/11/14 Type: Axial Exception Level: Sec/Soft Time: 03:35:45:600 Log Series: 223 mA out of 5 % or 2 mA tol Selected technic: 140 kV, 56 mA Tot. samples = 4 Tot. mA samples above tol. = 0, below tol. = 1 Avg. of mA samples above tol. = 0.0 mA, below tol. = 0.2 mA Avg. of tot. mA = 26.7 mA Error: 183183 kV out of 3 % tol See Long Description for error example Error: 183183 Suite: CT01 Host: OBC Proc: Generator File: gen tech.c 1.32 Method: No Method Line: 791 Function: XRAY GENERATION : KV Loop 960822 1230 10922 Scan: 0/0/3 Type: Static Exception Level: Sec/Soft Time: 12:17:46:610 Log Series: 529 kV out of 3 % tol Selected technic: 100 kV, 50 mA Tot. samples = 3 Tot. kV samples above tol. = 0, below tol. = 1 Avg. of kV samples above tol. = 0.000 kV, below tol. = 53.871 kV 960822 1230 10922 Avg. of tot. kV = 84.396 kV
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required