GE Healthcare · CTHiSpeed18311013379
HiSpeed Advantage RP/S
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 183110 KV fault detected: Cathode overcurrent See Long Description for complete error Wed Sep 28 08:31:12 2005 Error: 183110 Host: OBC Process: Generator File: gen_abort.c 1.31 Line: 266 Function: XRAY GENERATION : KV Loop Scan: 13379/2/2 Type: Helical Exception Level: Pri/Most Severe Time: 08:31:12:415 Log Series: 332 KV fault detected: Cathode overcurrent. Fault latch address = ffcff3 hex, Bit = D5
Applies to
- Models
- HiSpeed
- Product
- HiSpeed Advantage RP/S HISPEED ADVANTAGE RP SOFTWARE LFC TAPE Version: rp25.5 Created: Tue Jan 26 09:23:55 CST 1999 Copyright: (c) 1988, 1999 The General Electric Company HISPEED ADVANTAGE RP SOFTWARE LFC TAPE Version: rp25.5 Created: Tue Jan 26 09:23:55 CST 1999 Copyright: (c) 1988, 1999 The General Electric Company Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- D3101T / D3102T X-ray Tube Used In CT/i and HiSpeed Advantage Part #: D3102T Part Type: GP TUBES Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: TUBE HSA/CTI FRU: Yes Status: Orderable Comments: FLS - NEW 1994 TUBE CAT. FOR RELOAD REPLACES D1199ZR (RELOADED) .9 X .7FS, 1.2-1.2 FS HISPEED ADVANTAGE DIMS 37 X 27 X 28 (LWH) MAXIRAY EQUIVALENT IS D3109T TUBE FAN= 46-219465P1 AM SOURCE = GEMS-MM (MEXICO) *** FLS Tube Names Systems Compatible Heat Capacity Focal Spot Catalog # Model # Insert # Casing # Tube Unit # HSA(Zeus) HiSpeed CT/I, or HiSpeed Advantage 3.5MHu 0.9x0.7/1.2x1.2 D3102T MX165 46-309300G2 46-309500G2 46-309500G801 Tubes started shipping half machined, half cast casings in early 2005. This was stated to be a tempory change in process while a new vendor was being evaluated.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.