Revolution XR/d Wallstand 90 Degree calibration will not update on calibration summary screen See long descrip…

Revolution XR/d Wallstand 90 Degree calibration will not update on calibration summary screen

Symptom

Revolution XR/d Wallstand 90 Degree calibration will not update on calibration summary screen See long description FE was trying to complete the Wallstand 90 Degree cal but the date would not update on the calibration summary screen after the cal was complete. Looking closer at the summary screen, the Wallstand Stretcher Mode date was updating after running the Wallstand 90 Degree cal. If the Wallstand Stretcher Mode calibration was run, it would update it's own date properly. This site was an extended wallstand only configuration with the wallstand at the head end of the stationary rails. During the Wallstand 90 Degree cal, the gooseneck was at zero degrees and in the stretcher mode, the gooseneck was rotated 90 degrees off of zero.

Applies to

Models
Revolution
Product
Revolution XRd
Area
X-ray R&F

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.