LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Image recon is failing intermittently - document in RFS - critical issue See long description for details. Image recon is failing intermittently.
Error message
Error: 200352903 Reconstruction errors detected See Long Description for error. Fri Jul 29 20:27:00 2005 Error: 200352903 Host: whct Process: reconSelftest File: reconSelftest.c Line: 200 Reconstruction errors detected Had Recon Retry on image 414/4/1, the recon job did not succeed errors in 1 images during selftest --------------------------------------------------------------------------- Fri Nov 17 12:32:05 2006 Error: 200352903 Host: ct04 Process: reconSelftest File: reconSelftest.c Line: 201 Reconstruction errors detected Had [IG 1] image checksum mismatch errors in 4 images during selftest Error: 290012014 Warning: BP start failure See LD for details Mon Nov 5 15:36:50 2007 Error: 290012014 Host: ig3 Process: image generation File: Bp3dVRAC.cpp Line: 201 Warning: BP start failure Error: 290009007 Timeout waiting for image available. Number of retry so far: 1. IG node: 0. Sun May 14 14:05:42 2006 Error: 290009007 Host: darc Process: recon_control File: ImageSetRequest.cxx Line: 5208 Function: Image Generation : SBC Processing Job Type: RECON_JOB_PROS_HELICAL_RECON Scan: 378/2/2 Image: 378/2/39 Timeout waiting for image available. Number of retry so far: 1. IG node: 0.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5159834-3 Jarrell VIG with Improved EMC Shielding, replaces 5159834-2 see long description for details 5159834-3 Jarrell VIG with Improved EMC Shielding, replaces 5159834-2 Part #: 5159834-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Jarrell VCT IG with VRAC2, Improved EMC Performance FRU: Yes Status: Orderable Per ECR 2042316, 212008898, 5/8/07. ** FE Assist ** FE must order software release 06MW29.7, p/n 5193989 , via FEMC/Coakly and perform LFC to support this FRU hardware if system is not currently running sw rls 06MW29.7, or more recent ( 5/15/07 212001261)
- Area
- CT
Full cause & step-by-step resolution