Pular para o conteúdo principal
GE Healthcare · CT23000652925088

Lightspeed 1.X - Octane

Symptom

Image reconstruction failures. Image reconstruction failures

Error message

Error: 230006529 Recon Mgr is sending queue ack to queuemgr with a RECON_JOB_FAILED_RETRY status See Long Description for Full Error Tue Feb 1 06:33:34 2005 Error: 230006529 Host: ice Process: reconmgr File: rm_recon_events.c Line: 329 Function: Image Generation : SBC Processing Job Type: Prospective Axial Scan: 25088/2/9 Image: 25088/3/25 Recon Seq Id: 957 Recon Mgr is sending queue ack to queuemgr with a RECON_JOB_FAILED_RETRY status

Applies to

Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2225004-10 73 GB destroked to 18GB 15000 Used on Octane OC with 9Gb OS hard drive. See Long Description for more detail. This part is used on Octane Operators Consoles that have a 9Gb OS hard drive and need a destroked Scan Data Disk. See SN0711061b for jumper settings. Part #: 2225004-10 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: 73 GB destroked to 18GB 15000 RPM Scan Data Drive FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues