Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT29000900931809

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Center image artifact appears on head images and would clear by restarting the scanner See long description for artifact Center image artifact appears on head images and would clear by restarting the scanner

Error message

ig1 kernel: rac: PCI ERROR: Master Data Parity Error See LD for details ig1 kernel: rac: PCI ERROR: Master Data Parity Error rsh darc rsh ig1 ( ig1, ig2, or ig3 ) cd /var/log tail -1000 messages | more Look for any of the IG errors listed below: Aug 29 15:20:24 ig1 kernel: rac: PCI ERROR: Master Data Parity Error Aug 29 15:20:24 ig1 kernel: rac: PCI ERROR: Signaled Target Abort Aug 29 15:20:24 ig1 kernel: rac: PCI ERROR: Received Master Abort Aug 29 15:20:24 ig1 kernel: rac: PCI ERROR: Signaled System Error Aug 29 15:20:24 ig1 kernel: rac: PCI ERROR: Detected Parity Error Aug 29 15:20:24 ig1 kernel: rac: RAC reconfigure: current load -1; new load 0 Aug 29 15:20:26 ig1 kernel: rac: timed out reconfiguring device (1900 ms)

Do not Use - use one of the other Error: 290009009 Image failed due to many reasons See Long Description for complete error Tue Apr 19 12:44:49 2005 Error: 290009009 Host: darc Process: recon_control File: ImageSetRequest.cxx Line: 3968 Function: Image Generation : SBC Processing Job Type: RECON_JOB_PROS_AXIAL_RECON Scan: 4354/2/3 Image: 4354/2/6 Image failed due to: IG error: PROCESSING_ERROR. Number of retry so far: 1. Tue Nov 7 11:38:44 2006 Error: 290009009 Ho

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5159834-3 Jarrell VIG with Improved EMC Shielding, replaces 5159834-2 see long description for details 5159834-3 Jarrell VIG with Improved EMC Shielding, replaces 5159834-2 Part #: 5159834-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Jarrell VCT IG with VRAC2, Improved EMC Performance FRU: Yes Status: Orderable Per ECR 2042316, 212008898, 5/8/07. ** FE Assist ** FE must order software release 06MW29.7, p/n 5193989 , via FEMC/Coakly and perform LFC to support this FRU hardware if system is not currently running sw rls 06MW29.7, or more recent ( 5/15/07 212001261)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues