GE Healthcare · CTHiSpeed182904
HiSpeed CTi 6.X with Octane and HSA Gantry
Symptom
Gantry table controls would not work when moving patient into the gantry. Footpedals would move table up/down.
Error message
Error: 182904 A table side key is either stuck or pressed during initialization or just after Emergency Off released. See the Long Description for the full error code Thu Sep 16 06:48:08 2004 Error: 182904 Host: ETC Process: Table Controls File: tc_exec.c 1.27 1/15/96 Line: 345 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 06:48:00:070 Log Series: 20 A table side key is either stuck or pressed during initialization or just after Emergency Off released. Port value 0x0410. Disabling those keys detected being stuck
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required