GE Healthcare · CTLightSpeed260100096
LightSpeed 5.X (80/100kW)
Symptom
Exposure Interlock aborted scanning. Exposure Interlock aborted scanning
Error message
Error: 260100096 Exposure Interlock Opened See Long Description for complete error. 260100096 Exposure Interlock Opened Mon Dec 6 09:54:17 2004 Error: 260100096 Host: tgp Process: tgp_subsys File: scTrigSeqActions.c Line: 546 Mon Dec 6 09:54:17 2004 Host : Table Gantry Processor Ermes # : 260100096 Exception Class : Abort Severity : Primary File : scTrigSeqActions.c Line# : 546 Function : System Monitoring : Xray Exposure Interlocks Scan Type : Axial Scan: 7629/2/1 Exposure Interlock Opened. The current loop from Gantry to PDU. The cause of the failure: Console Relay
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required