LightSpeed 5.X (60kW)
Symptom
ECG Viewer is not receiving data from monitor when Gating BPM / ECG Trace button is clicked. Getting a pink pop-up box with ECG Viewer is not receiving data from monitor when they click on the Gating BPM / ECG Trace button with in the Fast Chest protocol. There are no errors posted in the messages box. . Getting a pink pop-up box with ECG Viewer is not receiving data from monitor when they click on the Gating BPM / ECG Trace button with in the Fast Chest protocol. There are no errors posted in the messages box. The information in the attention box is ECG Viewer is not receiving data from monitor. Viewer is closing. Please verify that the monitor is on and the ethernet cable is connected, then turn Gating OFF and ON to restart ECG Viewer
Error message
Error: 260100083 Timeout Occurred Waiting for an External Gate Signal. Check Gating Hardware/Patient Connection See Long Description for complete error Tue Jan 22 13:34:24 2008 Error: 260100083 Host: tgp Process: tgp_subsys File: scExternalGate.c Line: 1030 Tue Jan 22 13:34:23 2008 Host : Table Gantry Processor Ermes # : 260100083 Exception Class : Abort Severity : Primary File : scExternalGate.c Line# : 1030 Function : No System Function Reported Scan Type : Axial Scan: 1318/2/1 Scan Aborted. A Timeout Occurred Waiting for an External Gate Signal. Check Gating Hardware/Patient Connection.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5198566 cable-gantry - EKG PRO See Long Discription for Details Part #: 5198566 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: cable-gantry - exg pro FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required