GE Healthcare · CTLightSpeed260107205260107437
LightSpeed 5.X (80/100kW)
Symptom
"Scan hardware Stopped Scan" with Axial Errors. "Scan hardware Stopped Scan" with Axial Errors
Error message
Error: 260107205 Axial Motor Control (AMC) logic could NOT clear the FAULT line from the AXIAL Drive See Long Description Tue Jun 6 16:50:06 2006 Error: 260107205 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 2605 Tue Jun 6 16:50:07 2006 Host : Table Gantry Processor Ermes # : 260107205 Exception Class : Abort Severity : Primary File : amc_mgr.c Line# : 2605 Function : AXIAL CONTROL : Axial Communications Scan Type : None/Unknown Scan: 0/0/0 Axial Motor Control (AMC) logic could NOT clear the FAULT line from the AXIAL Drive. Last Axial Drive Fault will be logged before this message. Error: 260107437 The Axial AC Drive reported fault code: F:37 P Jump Error Fault see LD Error: 260107437 The Axial AC Drive reported fault code: F:37 P Jump Error Fault Tue Jun 6 16:50:07 2006 Error: 260107437 Host: tgp Process: tgp_subsys File: AX2fault.c Line: 195 Tue Jun 6 16:50:07 2006 Host : Table Gantry Processor Ermes # : 260107437 Exception Class : Abort Severity : Primary File : AX2fault.c Line# : 195 Function : AXIAL CONTROL : Axial Communications Scan Type : None/Unknown Scan: 0/0/0 %s The Axial AC Drive reported fault code: F:37 P Jump Error Fault
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2233402 Axial Drive Motor Controller See Long Description for details Part #: 2233402 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 8.64 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: AX DRIVE MTR CONTRLLER W/CORERS ASSY FRU: Yes Status: Orderable Comments: FLS - H2 GANTRY: MUST ORDER FROM HOLT ELECTRONICS. MS26561 DSM 1/7/01 USED ON HPOWER PROGRAM, PER RSPL. MS20544 ON 2/12/03. DO NOT DISCOUNT - PER SHAKUN*** 2ND OPINION NOTES *** Check fuses/connections associated with axial drive circuit; run axial functional diagnostic test and check for failures. If used on HP60
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.