Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT260006833260007323

LightSpeed 4.X - GRE

Symptom

"Scan hardware Stopped Scan" with Axial Errors. "Scan hardware Stopped Scan" with Axial Errors

Error message

Error: 260006833 Axial Motor Control (AMC) logic detected an Error with the Axial Drive LONG TERM Position Tracking. See LD for complete error Error: 260006833 Axial Motor Control (AMC) logic detected an Error with the Axial Drive LONG TERM Position Tracking. Sat Mar 25 12:52:11 2006 Error: 260006833 Host: STC Process: Axial File: amc_mon.c Line: 871 Function: AXIAL CONTROL : axial control Scan: 2592/2/1 Type: Axial Exception Level: Pri/Most Severe Time: 12:52:10:845 Log Series: 6 Axial Motor Control (AMC) logic detected an Error with the Axial Drive LONG TERM Position Tracking (TRIM) Error. Position TRIM Error: 2840. Position TRIM Tol: 1000. HW/SW Error: Retry Rx to verify probable Gantry Hardware problem.

Error: 260007323 Unacceptable Amount of Acquisition Trigger Jitter See LD for complete error Fri May 14 07:25:51 2004 Error: 260007323 Host: DAS Process: DCB Firmware File: abort.c Line: 315 Function: DATA ACQUISITION : Scan Data Trigger Generation Scan: 4371/1/1 Type: Axial Exception Level: Sec/Soft Time: 07:25:51:075 Log Series: 51 Scan ABORT -- DAS Has Detected An Error.^B Acquisition hardware (DCB) has detected an Unacceptable Amount of Acquisition Trigger Jitter. (Trigger 

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP, VCT, and HD750. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) See LD for details 5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP and VCT. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) Part #: 5311936 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Shipping collector for Axial Encoder Assembly. FRU: Yes Status: Orderable Comments: Per ECR 2058601, 212008898, 10/8/08
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.