GE Healthcare · CTLightSpeedVCT26000240339667
LightSpeed 2.X (Plus, QX/i) - GRE
Symptom
"Scan hardware Stopped Scan" with DAS power errors. "Scan hardware Stopped Scan" with DAS power errors
Error message
Error: 260002403 The Data Acquisition System Has Detected An Error In The -5V or +5V Analog DAS Power Supply. See Long Description for the full error code Tue Dec 7 07:20:51 2004 Error: 260002403 Host: DAS Process: DCB Firmware File: pwrtst.c Line: 783 Function: DATA ACQUSITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 07:20:50:240 Log Series: 23 The Data Acquisition System Has Detected An Error In The -5V Analog DAS Power Supply. Measured: 508 mV, Expected range: -5250 to -4750 mV Fri Jul 17 16:00:53 2009 Error: 260002403 Host: DAS Process: DCB Firmware File: pwrtst.c Line: 729 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 39667/9/1 Type: Helical Exception Level: Sec/Soft Time: 16:00:54:485 Log Series: 357 The Data Acquisition System Has Detected An Error In The +5V Analog DAS Power Supply.^B Measured: 4732 mV , Expected range: 4750 to 5250 mV
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 2.X (Plus, QX/i) - GRE always an SDAS 4 slice H2_P_S4 (SDAS4 and H2 gantry) Lightspeed 2.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 3.X and 4.X. Always has an SDAS 4 slice Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2225214 +/- 5V Analog H2 SDAS power supply includes the mounting plate, input/ output harness and power supply - used on 2.x and QXi See Long Description for details Part #: 2225214 Part Type: GP CONSUMABLES Part Weight: 5.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: POWER SUPPLY ASM CC CASE FRU: Yes Status: Orderable Parts Offered as: Current: 2225214 Harvested: Not Available Exchange: Not Available Comments: CT LIGHTSPEED H2 GANTRY 5V 6A ANALOG SUPPLY If no 2225214 power supplies are available part number 2218158 (H1 Gantry S-DAS) can be used, but the mounting bracket must be reused off the old power supply.
- Area
- CT
Full cause & step-by-step resolution