Pular para o conteúdo principal
GE Healthcare · CTLightSpeed182405100642072920807184836

LightSpeed 2.X (Plus, QX/i) - Octane

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 182405 HW/SW error: Uncommanded elevation motion. See long description for full error Wed Mar 19 13:22:41 2008 Error: 182405 Host: ETC Process: Elevation File: unComMotion.c 1.28 11/5/92 Line: 170 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 10064/2/6 Type: Axial Exception Level: Pri/Soft Time: 13:22:42:215 Log Series: 54 HW/SW error: Uncommanded elevation motion. Current Position 20729 (.01 mm) Last Stationary Position 20807 (.01 mm) Position Tolerance 75 (.01 mm) Current Encoder cnts 3446 Last Stationary Encoder Cnts 3435

Error: 184836 motor controller detects uncommanded motion. Aborting any active scan Tue Aug 10 19:04:09 2004 Error: 184836 Host: ETC Process: Elevation File: elev scan.c 1.17 9/30/92 Line: 178 Function: No System Function Reported Scan: 4530/2/7 Type: Helical Exception Level: Pri/Most Severe Time: 19:04:08:905 Log Series: 376 motor controller detects uncommanded motion. Aborting any active scan

Applies to

Models
LightSpeed
Product
LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-296854P2 Encoder - used on HSA/CTi/1.X gantries as the Axial Encoder and on Zeus Tables as the Elevation Encoder. See Long Description for part details Part #: 46-296854P2 Part Type: GP CONSUMABLES Part Weight: 0.88 LBR Repairable: No Returnable: No Hazardous Material: N Description: ELEVATION ENCODER FRU: Yes Status: Orderable ICV/Transfer Price: 104.55 USD /Each List Price: 815.40 USD /EACH Comments: FLS - ADDED PER PCN 189960. MS4636 4/22/97 & PCN 184587 MS03330 ** FE ASSIST ** Axial encoder HSA/CTi/1.X mounts on Axial Motor. (212001261 8/5/08) Part #: 46-296854P2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Axial Encoder FRU: Yes Status: Orderable Parts Offered as: Current: 46-296854P2 Used: Not Available Exchange: Not Available Comments: FLS - ADDED PER PCN 189960. MS4636 4/22/97 & PCN 184587 MS03330 5/5/97 COLLECTION OF DEFECTIVES START7/23/01 MS5277 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA FE ASSIST: IN ORDER TO INCREASE REVENUE GLOBAL PARTS HAS CREATED A PROGRAM TO HARVEST PARTS FROM RENEWABLE RESOURCES. THESE PARTS OFFER DIRECT SAVINGS TO THE BUSINESSAND THE FIELD. ORDER THE "U" PART IF AVAILABLE. IF NO "U" USED PARTS ARE AVAILABLE CARES WILL CHANGE THE ORDER TOA NEW PART.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.