Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT18212650091491163490966178517244555

LightSpeed 1.X - GRE

Symptom

Scanner Hardware Reset Failed. See LD Scanner Hardware Reset Failed Hardware Failed to reset message appeared on monitor AFTER boot-up WITH Error:178568 Failed to establish socket connection to STC in 10 seconds and errors 244553 and 244554 and power up self test fails A, B, and 9 on STC CPU board The following errors were in the gesyslog Thu Jan 13 04:18:32 2005 Error: 178568 Suite: RP25 Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to STC in 10 seconds Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset. Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC. Also saw failures A, B, and 9 on the power-up test on the STC CPU .

Error message

Error: 182126 LAN Watchdog TIMEOUT with OBC (uuid 193) See LD for complete error Wed Sep 20 08:39:07 2006 Error: 182126 Host: STC Process: Master Exc Handler File: exm_lan_wdog.c Line: 851 Function: No System Function Reported Scan: 50091/8/0 Type: None/Unknown Exception Level: Secondary Time: 08:39:01:815 Log Series: 724 LAN Watchdog TIMEOUT with OBC (uuid 193) (SW opening abort relay). HW/SW Error: Did not receive LAN Watchdog message from OBC. (Last Wdog Probe Sent:491163 Last Response Rcvd:490966 (in Ticks)).

Error: 178517 Failure writing to socket connected to controller OBCR. No route to host See LD for complete error Fri Oct 8 07:49:13 2004 178517 4 ct03 fwmgr NBSSocket.c 690 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failure writing to socket connected to controller OBCR. OS error : 113 --- No route to host May also see the following OS error: OS error : 101 --- Network is unreachable

Error: 244555 The OBC controller has failed to reset See Long Description for the full error Error: 244555 Host: ct01_sbc Process: fwmgr File: HwSetUp.c Line: 724 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq 

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 1.X - GRE H1_P_S4 Lightspeed 1.X - GRE Method of visual identification Large square gantry similar in appearance to a CTi gantry. Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2126034 EMI COMPLIANT RCOM see long description 2126034 EMI COMPLIANT RCOM 2126034 EMI COMPLIANT RCOM Part #: 2126034 Part Type: GP REPAIRABLES Part Weight: 2.27 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: EMI COMPLIANT RCOM ASSEMBLY FRU: Yes Status: Orderable Comments: FLS - HQ INFO: POOL TO BE OBTAINED BY UPGRADE OF T3202DY PER PCN 189224. REWORK ALL RG'S ASTHEY CYCLE THRU DBR. MS8365 MBELL 3/95 46-297474G1 REPLACED BY 2126034/T3202GK PER PCN 189275. MS4636 4/3/95 GELS - CAN REPLACE 46-297474G1. 46-297474G1 CAN NOT REPLACE 2126034. SP3/95 KANBAN CSS F00,EMI COMPLIANT RCOM ASSEMBLY This board is within the OBC
Area
CT