Pular para o conteúdo principal
GE Healthcare · CTHiSpeed183150

HiSpeed CTi 6.X with Octane and HSA Gantry

Symptom

Scanner Hardware Fault with high voltage spits. Scanner Hardware Fault with high voltage spits

Error message

Error: 183150 Rotor current out of 25 % WARNING tolerance See long description for full error Tue Dec 2 15:39:06 2008 Error: 183150 Suite: CT04 Host: OBC Process: Generator File: ctvrc manager.c 1.8 Method: No Method Line: 303 Function: XRAY GENERATION : Tube Rotor Control Scan: 5737/5/5 Type: Axial Exception Level: Pri/Soft Time: 15:39:06:810 Log Series: 905 Rotor current out of 25 % WARNING tolerance. Green stator = 3.974 A, Black stator =2.277 A. Reference current = 4.000 A. Speed = 2. (0 = low, 1 = medium, 2 = high) State = 1. (0 = accelerate, 1 = run)

Applies to

Models
HiSpeed
Product
HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
D3101T / D3102T X-ray Tube Used In CT/i and HiSpeed Advantage Part #: D3102T Part Type: GP TUBES Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: TUBE HSA/CTI FRU: Yes Status: Orderable Comments: FLS - NEW 1994 TUBE CAT. FOR RELOAD REPLACES D1199ZR (RELOADED) .9 X .7FS, 1.2-1.2 FS HISPEED ADVANTAGE DIMS 37 X 27 X 28 (LWH) MAXIRAY EQUIVALENT IS D3109T TUBE FAN= 46-219465P1 AM SOURCE = GEMS-MM (MEXICO) *** FLS Tube Names Systems Compatible Heat Capacity Focal Spot Catalog # Model # Insert # Casing # Tube Unit # HSA(Zeus) HiSpeed CT/I, or HiSpeed Advantage 3.5MHu 0.9x0.7/1.2x1.2 D3102T MX165 46-309300G2 46-309500G2 46-309500G801 Tubes started shipping half machined, half cast casings in early 2005. This was stated to be a tempory change in process while a new vendor was being evaluated.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues