GE Healthcare · CTHiSpeed183179
HiSpeed CTi 6.X with Octane and HSA Gantry
Symptom
Scan abort during any scan with external x-ray interlocks not satisfied.
Error message
Error: 183179 I/O board sensed external x-ray interlocks not satisfied. Sun Sep 5 11:10:28 2004 Error: 183179 Host: OBC Process: Generator File: gen_control.c 1.39 Line: 658 Function: XRAY GENERATION : Xray Exposure Interlocks Scan: 4167/1/1 Type: Scout Exception Level: Pri/Soft Time: 11:10:29:165 Log Series: 41 I/O board sensed external x-ray interlocks not satisfied. Possible causes: - Scanning room door is open (if applicable). - Check relay on DIP and/or Axial board. This error is flagged by the OBC on I/O board during xray generation.
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required